Error codes

Any malfunctions during operation will be identified by an error message on the display (E and F).
Proceed as follows in such cases:
Switch off device using the main switch (A)
Carry out corrective measures
Restart the device
Error code
Err
Err
Err
Err
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
092007
All manuals and user guides at all-guides.com
Cause
Undervoltage
Mains voltage failed during operation.
Mains plug connected with switch set to
"On".

Overtemperature

Device will switch automatically to cooling Motor speed - Allow device to cool down:
mode if operated with excess load
Electronic
Electronic fault
Effect
Solution
Device off
- Switch device off and on
Device off
- Switch device off and on
set auto-
- Allow to continue running in cooling mode
matically
- Switch device off. Switch on again after some time.
(not adjust-
(Duration depends on the magnitude of the overtempertature)
able)
A standard power meter is used between the mains socket
and the mains plug to prevent the device failing as a result
of excess load.
Power consumption < 500 W : continuous operation
Power consumption > 500 W : device may switch to cooling mode
Motor speed - Switch off device
set auto-
Please send the device to the Servicing department for repair,
matically (not enclosing a short description of the fault.
adjustable)