The fault is shown by an error code in the LED display (C) as following if the error occurs.
Proceed as follows in such cases:
F Switch off the device using the main switch
F Carry out corrective measures
F Restart the instrument
Error code Cause
– – –
The cap is not closed properly
E3
Temperature inside instrument is too high motor off
E4
Motor blockage or overload
E25
Safety relay fault
E86
Voltage is too low
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.
A10.1 Standard beater
A10.3 Hard metal cutter
In accordance with IKA
warranty conditions, the war-
®
ranty period is 24 months. For claims under the warranty
please contact your local dealer. You may also send the
machine direct to our factory, enclosing the delivery in-
voice and giving reasons for the claim. You will be liable
for freight costs.

Error codes

Effect
Solution
instrument
- Close the cover properly
does not start
- Switch off the instrument and allow it to
cool down
motor off
- Switch off the instrument
- Decrease the motor's load and reduce the
material to be ground
motor off
- Send the instrument for repair
motor off
- Switch off the instrument
- Check the Mains voltage

Accessories

A10.2 Star-shaped knife
A10.4 Grinding chamber reduction

Warranty

The warranty does not cover worn out parts, nor does
it apply to faults resulting from improper use, insufficient
care or maintenance not carried out in accordance with
the instructions in this operating manual.