Transport and storage

Check packaging and sensor for damage.
f
In the event of damage: Do not use sensor.
f
5.

Mounting

5.1 Mounting conditions
Fig. 3. Mounting locations
Fig. 4. Fully submerged in medium
Fig. 5. Mounting in flow direction
≥1 m
Fig. 6. Recommended distance from bend
www.baumer.com
CombiLyz AFI4/AFI5
1
Operating Instructions
Conductivity measurement
Inductive conductivity transmitter
Store sensor where it will be secure against shock.
f
Storage temperature: –30 ... 80 °C
Relative humidity: < 98 %
The sensor can be mounted at any point on the
vessel or pipe.
For correct operation the sensor must be fully
submerged in the medium.
To allow a sufficient self-cleaning to happen, the
hole through the sensor must be oriented in the
flow direction. Arrows (1) on the sensor above the
connection mark the orientation of the hole.
To avoid problems due to turbulence in flowing
media, Baumer recommends to mount the sensor
in a distance of at least 1 m from a bend.