Inductive conductivity transmitter
9.

Operation

9.1 Display views
Selectable display views
Conductivity/
concentration and tag
Conductivity/
concentration bar
Visual alerts and colors
White background

10. Troubleshooting

Fault
Display is off and no
signals from transmitter
Display is on but no signals
from transmitter
Display is off but there are
signals from transmitter
Display does not show the
correct data
14 / 68
Conductivity measurement
Conductivity and details
Conductivity/
concentration and
temperature bar
Green background
Cause
Sensor not correctly connected
Device error
Short circuit
UnitCom cable not connected
Unsuitable media characteristics
Operating Instructions
Media label and tag
Conductivity and
concentration/
temperature
Red background
Action
Check plug and power supply.
f
Dismount and return sensor.
f
Remedy short circuit.
f
Connect the UnitCom cable between
f
display and transmitter.
Check signal quality with
f
FlexProgrammer 9701.
Medial label and details
Concentration and
details
Red background and
error indication
www.baumer.com