CombiLyz AFI4/AFI5
Inductive conductivity transmitter
Connect the cable plug with the first calibration point in the measuring range (see box mark).
f
-
Enter the measured resistance value according to the measuring record of the calibration box in the text
f
field.
Click on Calibrate.
f
Repeat the steps for all measuring points that are to be measured. Connect the cable plug with the
f
corresponding resistor.
Press button Finalize calibration.
f
Result:
Calibration of measuring range completed.
NOTICE
This calibration method does not consider any change in the sensor geometry (e.g. by abrasion). If the
sensor tip changes dimensions, the measured conductivity value will be different from the previous one
measured with the original tip geometry.
Any change in geometry can be compensated with the installation factor.
For further information, refer to the HELP menu in FlexProgram.
12.4 Reset user calibration
NOTICE
This task can be done either using the FlexProgram or via IO-Link (e.g. BSS). The following example shows
calibration with FlexProgram.
Instruction:
Open software FlexProgram on the PC.
f
Open setting AFI in the left column.
f
Click on icon
Measuring (Online).
f
Click on button Reset User Calibration below the visualization.
f
Result:
The calibration is reset to the factory state. Offsets are set to 0.
18 / 68