CombiLyz AFI4/AFI5
Inductive conductivity transmitter
Temperature performance characteristics
Temperature
Measuring range
Step response time,
T90
Max. measuring error
Reference conditions
for max. measuring
error
Temperature coefficient
(factor of change in
ambient temperature of
25 °C)
Process conditions
Process temperature
Process pressure
SIP/CIP compability
Factory settings
Output mode
Conductivity range 1
Conductivity range 2
Conductivity range 3
Conductivity range 4
20 / 68
Conductivity measurement
Free programmable
range
–20 ... 150 °C
≤ 15 s
± 0.4 K
Sensor incl. transmitter
at 25 °C ambient tempe-
rature
≤ 0.05 K/K
–20 ... 140 °C,
permanent
140 ... 150 °C,
max. t < 1 h
≤ 25 bar
< 60 min, at medium
temperature up to 150 °C
Conductivity
0 ... 200 mS/cm
0 ... 20 mS/cm
0 ... 2 mS/cm
0 ... 500 μS/cm
Operating Instructions
Ambient conditions
Operating
temperature
Degree of protection
(EN 60529)
Humidity
Insulation voltage
Vibration (sinusoidal)
(EN 60068-2-6)
Output signal
Conductivity/
concentration
Temperature
Relays
Current rating
Interface
Power supply
Voltage supply
Power-up time
Temperature output
Output damping
Temperature com-
pensation range 1–4
Lower output current
limit
Upper output current
limit
–30 ... 80 °C, with DFON
touch screen
–40 ... 85 °C, without DFON
touch screen
IP67
IP69K, with appropriate
cable
< 98 % RH, condensing
500 V AC
1.0 mm p-p (2 ...13.2 Hz),
0.7 g (13.2 ... 100 Hz),
1 octave / min.
4 ... 20 mA
4 ... 20 mA + HART®
4 ... 20 mA
2 relays included in the
display
100 mA, max.
IO-Link 1.1
With HART® modem
With FlexProgrammer 9701
15 ... 35 V DC
18 ... 30 V DC, with IO-Link
≤ 10 s, without DFON touch
screen
≤ 16 s, with DFON touch
screen
0 ... 150 °C
0.0 s
2.0 % FSR/K
3.7 mA
21.0 mA
www.baumer.com