Mounting flange (301048)
Assembly:
- Drill a bushing bore with 10 ... 12 mm diameter in the wall.
- Align and drill the bore pattern for the fastening screws of the mounting
flange (301048) or the wall mounting flange (520181).
- Screw down the flange.
- Remove protective cap and insert sensor probe carefully in coaxial di-
rection into flange.
- Adjust immersion depth of probe and fasten the sensor by means of the
locking screw (mounting flange 301046) or lock nut (wall mounting flange
520181).
Mounting with compression fitting
SCHMIDT Technology offers two compression fittings differing in material
(brass or stainless steel).
MID = 58.5
9,3
- The compression fitting is mounted using its external thread G½.
- Normally, a clamp with internal thread (G½ or Rp½) is welded as a con-
necting piece onto the bore of the wall and the fitting is screwed in.
- The further assembly is carried out as described previously.
Mounting accessories are available from SCHMIDT
www.schmidttechnology.de/en/produkt/schmidt-flow-sensor-ss-20-225/
Gebrauchsanweisung / Instructions for Use - SCHMIDT
5,8
SL
32
Wall mounting flange (520181)
SL
Length of welded clamp [mm]
MID Minimum immersion depth [mm]
D
Outer diameter of pipe [mm]
O
®
at:
®
Flow Sensor SS 20.225