Rear panel
See "B" on the front fold-out page.
1
AUX input (AUX)
The AUX input is currently not
supported.
2
Analog input (ANALOG IN)
3
Ethernet connection for connecting a PC
(LAN)
4
Connecting screw between GND and PE
The connecting screw must not be
removed!
5
Power switch
6
Power connection
7
Line fuse
8
Single-ended synchronization input for CLK
signal (SYNC.IN CLK)
9
Single-ended synchronization input for PTM
signal (SYNC.IN PTM)
10
Single-ended synchronization output for CLK
signal (SYNC.OUT CLK)
11
Single-ended synchronization output for PTM
signal (SYNC.OUT PTM)
12
Position Data Link (PDL)
13
Differential synchronization input for CLK or
PTM signal (SYNC.IN)
14
Differential synchronization output for CLK or
PTM signal (SYNC.OUT)
15
Digital input (DIGITAL IN)
16
Trigger input (TRIGGER IN)
17
Trigger output (TRIGGER OUT)
HEIDENHAIN | EIB 8000 | 09/2018
NOTICE
4.3
Connecting the line voltage
WARNING
Risk of electric shock!
Improper grounding of electrical devices may result in
serious personal injury or death by electric shock.
Always use 3-wire power cables
Make sure the ground wire is correctly connected to
the ground of the building's electrical installations
WARNING
Fire hazard due to wrong power cable!
Use of a power cable that does not meet the
requirements of the mounting location may cause a fire
hazard.
Use only a power cable that meets at least the
national requirements of the respective country in
which the product is mounted
Use the supplied power cable to connect the power
connection to a 3-wire grounded power outlet
For information on the location of the power connection,
see "Rear panel", Page 29.
For pin layout see "K", Page 153.
4.4
Connecting the encoders
NOTICE
Damage to the product and the encoder caused by
incorrect voltage supply range or incorrect wiring!
An incorrect voltage supply range or incorrect wiring /
pin layout may cause damage to the product and the
encoder.
Note the voltage supply range of the connected
encoder
Verify that the connecting cable between the
encoder and the product is correctly wired
Vacant pins or wires must not be used
Do not engage or disengage the connecting cable
between the encoder and the product while under
power
The connection of non-HEIDENHAIN encoders to
the product and the operation of the product with
non-HEIDENHAIN encoders are at the user's own
risk