Fault
Reason
Yellow LED off
No power or power too low
with uncovered
(green LED does not light up)
path
Emitter defective
Path of rays covered (e.g. win-
dow dirty)
Wrong installation
Wrong settings
Yellow LED on
Receiver defective
with covered
Attenuation of the product too
path
low
Strong fluctu-
Turbulent application, reflections
ation of signal
and so on
strength
Endress+Hauser
Remedy
Check power supply
Replace emitter
Check path of rays and clean if
necessary
Check installation
11
È
Check settings
50
È
Replace receiver
Adjust sensitivity
53
È
Increase signal damping
63
È
en - Troubleshooting