Operation

A voltage between 12 V and 250 V can be connected at the contacts. The maximum
load must not exceed 8 A. The minimum power rating is 100 mA/5 V.
I
NOTE
The external alarm signal can only be switched off by eliminating the
cause of the alarm.
7
Operation
7.1
Before first use
Before first use, clean the refrigerator inside and outside with a damp cloth for
hygienic reasons (see also chapter "Cleaning and maintenance" on page 22).
7.2
Saving energy
• Only open the mother milk cooler as often and for as long as necessary.
• For optimal energy consumption, position the shelves and drawers according to
their position on delivery.
• On a regularly basis, make sure the door seal still fits properly.
• Clean dust and dirt from the condenser at regular intervals.
7.3
Switching on
!
CAUTION!
Do not store mother milk before the E6 error code has disappeared and
the operating temperature range has been reached.
• At the first starting up of the refrigerator or after switching it off for a long time the
A
display of the digital thermometer (fig. 2, page 3) shows the current ambient
temperature alternating with the error message „E6".
• The alarm LED will light up but no audible alarm will sound.
• As soon as the inner temperature reaches the operating temperature range
(1 °C to 4 °C) the current temperature is shown.
1. Reset the thermometer to avoid a wrong maximum temperature value.
2. Press the
button for 5 seconds.