13. ELECTROMAGNETIC COMPATIBILITY

The Swiss LithoClast® Trilogy should not be used adjacent to or stacked with another unit. If adjacent or
stacked use is necessary, the Swiss LithoClast® Trilogy should be observed to verify normal operation in the
Portable and mobile RF communications equipment should be used no closer than 30 cm to any part of the
Swiss LithoClast® Trilogy, including cables.
The Swiss LithoClast Trilogy uses RFID transmission: 13.56MHz, 1W RF output power, to communicate with probe.
Instructions for maintaining BASIC Safety and Essential performance for the expected Service life
The presence of transmitters near the Swiss LithoClast® Trilogy could affect its performances. The distances mentioned
in the tables prepared by manufacturer could help to prevent any disturbances of the equipment in normal operation.
The climatic environmental conditions could affect the life of critical components of the Swiss LithoClast® Trilogy.
Guidance and manufacturer's declaration – electromagnetic emissions
The Swiss LithoClast®
customer or the user of the Swiss LithoClast® Trilogy should assure that it is used in such an environment.
EMISSIONS TEST
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
emissions IEC
IEC 61000-3-3
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
The Swiss LithoClast® Trilogy uses RF energy only for its internal
Group 1
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
Class A
The emissions characteristics of the Swiss LithoClast® Trilogy
make it suitable for use in industrial areas and hospitals (CISPR 11
Class A
class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might
not offer adequate protection to radio-frequency communication
Complies
services. The user might need to take mitigation measures, such
as relocating or re-orienting the equipment.