Guidance and manufacturer's declaration – electromagnetic immunity
The Swiss LithoClast®
customer or the user of the Swiss LithoClast® Trilogy should ensure that it is used in such an environment.
IMMUNITY TEST
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 2 kV, 100 KHz for power
Electrical fast transient /
burst IEC
±1 kV, 100 KHz for input/
61000-4-4
± 0.5, ± 1 kV line(s) to line(s)
Surge
±0.5, ±1, ± 2 kV line(s) to
IEC 61000-4-5
Power frequency
(50/60 Hz) magnetic
IEC 61000-4-8
<5 % UT (>95 % dip in UT) for
40 % UT (60 % dip in UT) for
70 % UT (30 % dip in UT) for
Voltage dips, short
interruptions and
0 % UT for 0,5 cycle at 0°,
voltage variations on
45°, 90°, 135°, 180°, 225°,
power supply input lines
IEC 61000-4-11
0 % UT for 1 cycle single
<5 % UT (>95 % dip in UT)
38
IEC 60601
TEST LEVEL
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
supply lines
output lines
earth
30 A/m
<5 % UT (>95 % dip in UT) for
0,5 cycle
40 % UT (60 % dip in UT) for
5 cycles
70 % UT (30 % dip in UT) for
25 cycles
270° and 315°
phase
for 5 s
0% UT for 250 cycles
Table 9
COMPLIANCE LEVEL
± 2, ± 4, ± 6, ± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 2 kV, 100 KHz for power
supply lines
±1 kV, 100 KHz for input/
output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
30 A/m
0,5 cycle
5 cycles
25 cycles
0 % UT for 0,5 cycle at 0°,
45°, 90°, 135°, 180°, 225°,
270° and 315°
0 % UT for 1 cycle single
phase
<5 % UT (>95 % dip in UT)
for 5 s
0% UT for 250 cycles
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Floors should be wood, concrete
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial.
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Swiss LithoClast® Trilogy
requires continued operation
during power mains interruptions,
it is recommended that the Swiss
LithoClast® Trilogy be powered
from an uninterruptible power
supply or a battery.
UT is the a.c. mains voltage
(100-240) prior to application of
the test level.