Press the
(Help) key to see the on-screen instructions for
additional information.
4. Apply a new BiliCal tip to the hand-held unit. Firmly press it on
to the unit to ensure proper seating.
5. Select Done to begin calibration. The unit will beep and the
display will indicate Calibration successful.
Note: The unit will not beep if the sound is set to level 0.
6. Pull on the BiliCal tab and peel away the protective covering and
calibration material from the tip and dispose, leaving only the
clear film on the BiliCal tip. Press the
Done.
7. Gently press the BiliCal tip against the patient's forehead or
sternum. Lift the BiliCal tip from the patient's forehead or
sternum and then repeat the process four more times.
8. A single tone sounds after each reading to indicate that it was
successful. If an error occurs during the reading, a yellow error
screen will appear with a description of the error. Follow the
on-screen troubleshooting instructions to correct the problem.
Note: The unit will not beep if the sound is set to level 0.
10. When the readings are complete, the BiliChek displays and saves
all entered information.
11. Remove and dispose of the BiliCal tip and return the hand-held
unit to the charger base. If your unit is linked to a laboratory
information system, all new data entered since the last transfer
will be sent once the unit is placed back on the charger base.
(Enter) key to select
Chapter 4. Viewing and Changing Settings
WARNING
It is critical to calibrate the unit
and apply a new BiliCal Individual
Calibration Tip immediately prior
to performing a test. Inaccurate
test results will occur if this is not
done. Removal of the tip after
calibration and prior to a patient
measurement will result in loss of
calibration. The entire calibration
sequence must be repeated with a
new BiliCal Tip.
Note: The BiliChek unit should
always be stored on the charger
base when it is not in use.
TcB Records Sent Screen
23