6. Commissioning, operation

6.3.6 CALIBRATION mode (calibration of a pressure gauge)

1. Test point
(e.g. 0 bar)
EN
Deviation
No. of test item
and No. of test
point
2. Test point
(e.g. 1 bar)
Deviation
No. of test item
and No. of test
point
2. Test point
Deviation
No. of test item
and No. of test
point
x. Test point
42
All manuals and user guides at all-guides.com
Generate the specified set point
in accordance with the test
item display using the pressure
generator
If the set point = 0 bar, the
calibration assembly must
be brought to a pressure-free
condition/vented to atmosphere
(test item must indicate 0 bar;
ENTER
if necessary, make a zero point
setting) and with
Record the measured values of
the test point
Generate the specified set point
in accordance with the test
item display using the pressure
generator
Record the measured values of
ENTER
the test point
(with
test point)
WIKA operating instructions, model CPH7650