INSTALLATION

CONDITIONS FOR USE

• Compressors may only be installed and operating in dry, well-ventilated and clean environments
with environmental parameters that meet the requirements specified in Chapter 7, Technical data.
The compressor must be installed so that it is accessible at all times for operating and
maintenance. Please ensure that the label on the device is readily accessible.
• The compressor must stand on a flat, sufficiently stable base (be aware of the weight of the
compressor, see Chapter 7, Technical data).
• Compressors may not be operated outdoors or in otherwise wet or damp environments. Do not
use the compressor in the presence of explosive gases, dust or flammable liquids.
• Before connecting the compressor to medical equipment, the supplier must confirm that it meets
all requirements for its use. Refer to the technical data of the product for this purpose. When a
unit is to be built-in, classification and evaluation of compatibility must be done by the
manufacturer or supplier of the product to be used.
• Any other use or use outside this framework is not considered as intended use. The manufacturer
is not responsible for any damages arising from such use. The operator/user assumes all risks.
Only a qualified professional can install the compressor and place it into
operation for the first time. This professional is obliged to train operating staff as
to the use and maintenance of the device. Installation and training of all operators
shall be confirmed by the installer's signature on the certificate of installation.
Prior to installation, ensure that the compressor is free of all transport packaging
and stabilizers to avoid any risk of damage to the product.
Parts of the air pump may be hot and reach hazardous temperatures during
compressor operation and may pose a contact risk for operators or materials.
Burn or fire hazard! Caution! Hot surface!
You may notice a "new product" odour when you first place the product into
service (for a short period of time). This odour is temporary and does not impede
the normal use of the product. Ensure the space is properly ventilated after
installation.
Ambient operating conditions
Temperature:
Max. relative humidity:
Max. absolute humidity:
NP-DK50 PLUS, 2V-14_07-2021-MD
+5°C to +40°C,
70%,
3
15 g/m
- 26 -
DK50 PLUS, DK50 2V
07/2021