Tools List; T-Mass I 300, 500-Digital - Endress+Hauser t-mass I 300-digital Einbauanleitung

Replacing the sensor guard, pipe union and seal
Vorschau ausblenden Andere Handbücher für t-mass I 300-digital:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EA01290D
6

Tools list

Circlip pliers
7

t-mass I 300, 500-digital

7.1
Removing the sensor I
WARNING
L
The measuring device is energized!
Risk of fatal injury from electric shock.
Open the measuring device only when the device is de-energized.
CAUTION
L
Risk of injury!
To remove the sensor, the pipe must be completely free from pressure. Residual pressure
on the pipe can cause the sensor to shoot out as soon as the coupling is loosened!
In the case of toxic, explosive or flammable fluids, the pipe in which the measuring device
is installed must be rinsed or cleaned before the sensor is removed!
Before starting work, allow the pipe to cool down to a safe temperature.
1 →
32 mm/
36 mm/
41 mm
Loosen the coupling.
8
Hexagon wrench
32 mm, 36 mm,
41 mm
2 →
t-mass 300: Remove the
measuring device.
Allen key
3 mm
3
t-mass 500-digital: Remove the
measuring device.