Tools list
Circlip pliers
7
t-mass I 300, 500-digital
7.1
Removing the sensor I
WARNING
L
The measuring device is energized!
Risk of fatal injury from electric shock.
‣
Open the measuring device only when the device is de-energized.
CAUTION
L
Risk of injury!
‣
To remove the sensor, the pipe must be completely free from pressure. Residual pressure
on the pipe can cause the sensor to shoot out as soon as the coupling is loosened!
‣
In the case of toxic, explosive or flammable fluids, the pipe in which the measuring device
is installed must be rinsed or cleaned before the sensor is removed!
‣
Before starting work, allow the pipe to cool down to a safe temperature.
1 →
32 mm/
36 mm/
41 mm
‣
Loosen the coupling.
8
Hexagon wrench
32 mm, 36 mm,
41 mm
2 →
‣
t-mass 300: Remove the
measuring device.
Allen key
3 mm
3
‣
t-mass 500-digital: Remove the
measuring device.
Endress+Hauser