•Calibration
General
The UCV/R is factory calibrated. Once calibrated it will operate with any
input format.
Allow the instrument 1 minutes of powered operation (to reach a
stable temperature) before calibration.
During calibration allow a few seconds for signals to stabilise before
accepting the value.

Equipment requirements

• A suitable current or voltage source
• An accurate digital multimeter (accurate to 0.0 mV and ±0.1 μA)
Terminalconnectionsforinputcalibration
CalibrationStage
Current (mA) inputs
Voltage (V) inputs
Signaltype
Millivolt source +
Millivolt source −
Voltage source +
Voltage source −
Terminal
2
3
4
3
21