Herunterladen Diese Seite drucken

Stryker Surgistool 0830 Bedienungsanleitung Seite 17

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 40
D D i i s s i i n n f f e e c c t t i i n n g g t t h h e e p p r r o o d d u u c c t t
N N o o t t e e - - Always make sure that you wipe down each product with clean water and dry each product after disinfecting. Some
disinfecting products are corrosive in nature and may cause damage to the product if you use them improperly. If you do not
properly rinse and dry the product, a corrosive residue may be left on the surface of the product that could cause premature
corrosion of critical components. Failure to follow these disinfecting instructions may void your warranty.
Recommended disinfectants:
Quaternaries (active ingredient - ammonium chloride) that contain less than 3% glycol ether
Phenolic disinfectant (active ingredient - o-phenylphenol)
Isopropyl alcohol (active ingredient less than or equal to 21%)
Recommended disinfecting method:
1. Follow the disinfectant solution manufacturer's dilution recommendations.
2. Hand wash all surfaces of the product with a disinfectant solution.
3. Avoid oversaturation and make sure the product does not stay wet longer than the chemical manufacturer's guidelines
for proper disinfecting.
4. Dry the product.
5. Verify proper operation before you return the product to service.
0830-009-005 Rev AA.0
13
EN

Werbung

loading