Shut-down

Care

Stripping down

Disinfecting and cleaning

Shut-down
When therapy is finished,
Switch off the blue phototherapy lamps.
Switch off the white lights (where applicable).
Disconnect from the mains power supply.
Remove the phototherapy unit from the incubator /
warming cot canopy.
When the mobile stand has been used:
Release the castor brakes (where applicable).
Move the stand away from the incubator/warming cot.
Disconnect from the mains power supply.
Re-adjust the angle of lamp to the »zero« position.
Loosen the clamp and lower stand to transport height.
Store the phototherapy unit in a safe place.
Care
Stripping down
Lift the phototherapy lamp off the stand.
Wipe off any obvious stains with a disposable cloth.
Disinfecting and cleaning
Use only preparations classified as "surface disinfectants"
for disinfecting.
For material compatibility, we recommend preparations
based on
– aldehydes
– quaternary ammonium compounds.
Due to the chemical composition and direct effect on
material compatibility, preparations based on the
following components are of only limited suitability:
– halogen-releasing compounds
– strong organic acids
– oxygen-releasing compounds.
In general, the manufacturers' recommendations must be
followed when selecting the preparation. The manufacturer
is legally liable for the information concerning the range of
applications and possible material damage caused by
these preparations.
20
Do not use any disinfectants or cleaning agents
containing alcohol.
For users in the Federal Republic of Germany, we
recommend the use of disinfectants listed in the current
DGHM list (DGHM = German Society for Hygiene and
Microbiology).
The DGHM list (published by mhp Verlag GmbH,
Wiesbaden) also specifies the active basis of each
disinfectant. For countries where the DGHM list is
unavailable, we recommend products based on the
above active bases.
The following surface disinfectants can be recommended:
Dismozon
pur
Bode Chemie GmbH & Co,
Hamburg
Incidur
Henkel Hygiene GmbH,
Düsseldorf
Sekusept
pulver
Henkel Hygiene GmbH,
Düsseldorf
Follow the manufacturer's Instructions for Use.
Wipe-disinfecting the phototherapy lamp
Wipe off obvious stains using a disposable cloth and
detergent.
Wipe-disinfect surfaces.
After disinfecting, wipe surfaces again with a clean
damp cloth and then dry thoroughly.
Do not allow any liquid to get into the lamp.
Use only the recommended cleaning agents and disinfec-
tants, as there is a risk of tension cracks in the acrylic
glass if, for instance, alcohol is used.
After each care/cleaning/disinfecting operation,
check that the unit is functioning correctly.
(see page 19, "Switching on lamp")