In the standard specifications, the minimum phase difference is fixed at 50 ns, but the minimum
phase differences in the table below are available as special specifications.
A/B minimum
Phase A
phase difference
cycle
50 ns
200 ns
100 ns
400 ns
300 ns
1.2 µs
500 ns
2 µs
5-2. Output Signal Alarm
If the response speed is exceeded, the phase A/B output from this measuring unit becomes
high impedance for about 400 ms to serve as an alarm.
5-3. Receiver
Line driver
Output: phase A/B/Z
* If extending the cable, the supply voltage is +5 V±5%.
* Use the CE22 series extension cables for bare wires (optional accessories).
6
(E)
Counter
Maximum response
allowable
speed (Resolution
frequency
0.5 µm)
5 MHz
250 m/min
2.5 MHz
100 m/min
833 kHz
33 m/min
500 kHz
20 m/min
Phase A/B is high impedance
20 m or less
Remarks
Standard product
Special specifications
Special specifications
Special specifications
Alarm section
Receiver
Line receiver
AM26C32 or equivalent
DK50 / DK100 Series