Herunterladen Inhalt Inhalt Diese Seite drucken

Operation Procedure; Variosurg 3 Link Operation - NSK Wireless Link Tool Link Module LMP2 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 14
• If any abnormalities are found with the Surgic Pro2 and the VarioSurg 3, refer to the
respective operation manuals.
• If the button is not released after the LED turns blue, the LED blinks white and pairing
stops. To pair, release the button, and after the LED turns white, proceed from step 7.
• Pairing operation is required only for the first time. After that, the Link Module is
automatically connected when the Surgic Pro2 and the VarioSurg 3 are turned on. To
unpair the Surgic Pro2 and the Link Module, the Surgic Pro2 needs to be operated.
Refer to the Surgic Pro2 operation manual for details.
6
CAUTION
NOTICE

Operation Procedure

5
Turn on the paired Surgic Pro2 and the VarioSurg 3 connected to the Link Module.
5–1 VarioSurg 3 link operation
1
Make sure that the LED of the Link
Module blinks green and the Bluetooth
mark is on the left of VARIOSURG icon
on the Surgic Pro2 display.
2
Press and hold the LINK key of the
Surgic Pro2 for more than 2 seconds.
The Surgic Pro2 displays "LINK" and the
VarioSurg 3 can be operated with the
foot control.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis