Stryker TruRize 3333 Bedienungsanleitung Seite 25

Klinikstuhl

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 63
D D i i s s i i n n f f e e c c t t i i n n g g
C C A A U U T T I I O O N N
• Do not clean, disinfect, service, or perform maintenance while the product is in use.
• Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and motors.
Remove the occupant from the product, clean up the fluid, and inspect the product. Fluids can cause unpredictable
operation and decreased functionality of any electrical product. Do not return the product to service until it is completely
dry and you have tested for safe operation.
• Always wipe with clean water (or 70% isopropyl alcohol, if using V V i i r r e e x x ® TB) and dry each product after disinfecting.
Some disinfectants are corrosive in nature and may cause damage to the product. If you do not rinse and dry the
product, you may leave a corrosive residue on the surface of the product. This corrosive residue could cause premature
degradation of critical components. Failure to follow these disinfecting instructions may void your warranty.
The recommended disinfectants for this product's surfaces include the following:
Quaternary (active ingredient - ammonium chloride) that contain less than 3% glycol ether
Phenolic (active ingredient - o-phenylphenol)
Chlorinated bleach solution (use up to UK disinfecting 10,000 ppm available chlorine (941 mL of a 5.25% sodium
hypochlorite solution per 4000 mL of water))
Alcohol (active ingredient - 70% isopropyl alcohol)
≤ 21% isopropanol alcohol
To wipe down the product with disinfectant between uses:
1. Follow the manufacturer's dilution recommendations exactly.
2. Apply the recommended disinfectant solution by spray or pre-soaked wipes.
3. Hand wash all surfaces of the product with the recommended disinfectant.
4. Disinfect all exposed surfaces. Pay attention to high contact areas, such as the footrest cushion, seat cushion, backrest
cushion, footrest covers, armrest overmold and covers, and backrest mobility handles.
5. Follow the disinfecting solution manufacturer's instructions for appropriate contact time and rinsing requirements.
6. Dry the product thoroughly before returning the product to service.
Avoid over saturation. Do not allow the product to remain wet.
Follow the manufacturer's dilution recommendations for appropriate contact time and rinsing requirements. Follow the
chemical manufacturer's guidelines for proper disinfecting.
3333-009-005 Rev C.0
21
EN