N N a a m m e e
A A
Chair exit alarm
B B
Chair alarm arm/disarm
C C
Stand Assist
Upright 1
D D
Upright 2
E E
F F
Recline
G G
Full-Flat Transfer
H H
Trend
Chair position indicator
I I
Seat height up
J J
Seat height down
K K
Occupant control lock/unlock
L L
Parking brake status indicator
M M
Battery charging indicator (option)
N N
Battery status indicator (option)
O O
P P
Power indicator
Diagnostics indicator
Q Q
R R
Service indicator
3333-009-005 Rev C.0
F F u u n n c c t t i i o o n n
Illuminates green when you arm the chair exit system.
Flashes yellow and triggers an audible alarm when an
occupant exits the chair.
Press once to arm the chair exit system. Press again to
disarm the chair exit system.
The minimum occupant weight required to detect an
occupant to arm the chair exit system is 65 lb
Stand Assist, Full-Flat Transfer and Trend positions
are unavailable when you arm the chair exit system
The chair exit system will only arm when in the
Upright 1, Upright 2, and Recline positions and
when you apply the parking brake
The chair exit system alarm continues to sound until
you press arm/disarm
Provides lift assistance for occupants who may have
difficulty rising from a seated to a standing position.
Only available when you apply the parking brake.
Places the product into the Upright 1 position
Places the product into the Upright 2 position and
extends the footrest
Lowers the backrest and tilts the seat back
Places the product into a flat position for transfer to or
from a bed or stretcher
Places the product into the Trend position (head down
with foot up)
Follows the chair's position status from one position to
another
Raises the seat height
Lowers the seat height
Lock illuminates white when you lock the occupant
control panel. Unlock illuminates white when you
unlock the occupant control panel.
Lock illuminates white when you apply the parking
brake. Unlock flashes white when you release the
parking brake.
Illuminates to indicate that the battery is charging
Indicates the charge left in the battery
Illuminates when the product is powered
Illuminates when in diagnostics mode. Unplug product
and call maintenance.
Power the product off, then power the product back on
( Powering the product on and off (page 11)). If indicator
is off, continue normal operation. If indicator
illuminates, unplug product and call maintenance.