Herunterladen Inhalt Inhalt Diese Seite drucken

Acoustic Signal; Operating The As Super 4 Digital; Needle Placement And Connecting; Switching On The Device - schwa-medico AS SUPER 4 digital Gebrauchsanweisung

Elektrischer nadelstimulator
Vorschau ausblenden Andere Handbücher für AS SUPER 4 digital:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

30
schwa-medico
excess of 4 mA, and automatically discontinues the stimulation if:
»
a needle/electrodes detaches itself from the skin
»
needles/electrodes are not connected correctly in pairs
»
needles/electrodes are not sufficiently conductive
»
the connection between the cables and the needles/electrodes is interrupted
»
the connection between the cables and the device is interrupted
The intensity then returns to 0 mA; this is indicated by a short acoustic signal.

Acoustic signal

The AS super 4 digital emits an acoustic signal at each change
of settings, switching ON/OFF and at the end of a therapy
session.
To switch OFF the acoustic signal press simultaneously the
key of channel 1 and channel 3 until the "beep" menu
appears on the display. Press the
the key to get out of the acoustic signal mode.

Operating the AS SUPER 4 digital

Needle placement and connecting to the device
Connect the cables to the AS SUPER 4 digital unit. Place the needles into the skin on
the appropriate acupuncture points. Connect each needle with the crocodile clips at
the end of the cables. Please clip at the lower part of the needles in order to keep the
needles approximately straight. The crocodile clip should not be in contact with the
patient's skin. Please note that you always need 2 needles per cable.

Switching ON the device

Press the key to switch ON the AS SUPER 4 digital.

Program selection

Program group
Select a program group (1.0 to 9.0) by pressing the
To scroll through the program groups, press the
as necessary until you have selected the program group of
your choice. Once you have reached the last program group,
you will return to the first one.
Subprogram
Select a subprogram by pressing the
To scroll through the subprograms, press the
selected the subprogram of your choice. Once you have reached the last subprogram,
you will return to the first one.
Instructions for use - AS SUPER 4 digital
key to switch ON and OFF the acoustic signal, press
E
P
key.
P
key as often
key.
E
key as often as necessary until you have
E

Werbung

Kapitel

Inhaltsverzeichnis
loading