Herunterladen Inhalt Inhalt Diese Seite drucken

GCE MEDIEJECT II Bedienungsanleitung Seite 6

Absauggeräte
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 23
Free flow suction at inlet pressure
400 kPa:
Noise - close/open suction:
Outlet connection:
EN
* For gas used, see Annex No. 1
The parameters for the respective variant can be found in Annex No. 1.
According to EN ISO 10079-3 it is a device with High vacuum level at
least 60 kPa below atmospheric pressure and with High flowrate at least
20l/min.
1) Reference conditions: 20°C;1013hPa
2) The achievable vacuum depends on atmospheric pressure.
Every 100m increase in altitude causes a decrease in the achievable vacu-
um of about 1.4 kPa
3) Total accuracy of vacuum level is ± 8kPa
4) The control range of free suction flow is 10-25l/min
6. OPERATIONS
6.1. CONNECTION & USE OF INLET (PATIENT SIDE)
MediEject II is used with the optional associated equipment specified in
Chapter 8.
MediEject II installation must be in accordance with EN ISO 10079-3 and
EN 1789 as amended. For connection, see the figure below.
3.
1. Suction ejector
2. Filter
3. Suction hose
4. Collection container*
During suction, MediEject II must be used with one or more collecting
containers, which must include an overflow protection device to prevent
liquid from entering the suction ejector.
3.
4.
6/176
25 l/min
4)
35/45 dB
Hose nipple - inner diameter 6 mm
1.
2.
3.