7.6 PUTTING INTO OPERATION

If the cabinet has been sharply tilted while being sited or for mounting of accessories, it must be left to stand for
at least three hours before being started to allow the lubricating oil to settle in the compressor.
Failure to do so can cause total destruction of the compressor refrigerating system.
Check that the voltage on the identification plate is the same as that of the network.
THE ELECTRICAL SUPPLY LINE TO WHICH THE REFRIGERATOR WILL HAVE TO BE CONNECTED
IS TO BE PROTECTED BY A VERY SENSITIVE DIFFERENTIAL SWITCH (Rc=16A,Dc=30mA); IT IS
ALSO TO BE LINKED TO THE GENERAL EARTHING PLANT.
Earthing is necessary and compulsory by law.
Absolutely never use reduction units, patch cords, adapters or multiple socket outlets to connect
the counter. DIRECTLY connect the equipment to the mains.
The cable should be fully extended and located where it is unilikely to be struck, caught or stepped
on, away from water and the other liquids and heat sources and must not be damaged.
In case of breakage of the power cable of the cabinet, it must be replaced by the manufacturer or
by instruction it change of it.
Follow the manufacturer's instructions to install the unit, using qualified personnel and complying
with the electrical regulations for the specific country.
Incorrect installation may cause injury and damage; the manufacturer is not liable for this.
The manufacturer disclaims all responsability for any damage to people or things due to incorrect
observance of this rule.
7.7 LOADING
Warning! It's absolutely forbidden to place the product in the tanks over the max load level.
It is fundamental not to exceed the load limits indicated to avoid altering correct air circulation
and thereby avoid a higher product temperature.
Warning!
Keeping pharmaceuticals, glass bottles or flasks in the cabinet is strictly forbidden as they could break
and entail a safety hazard.
Attention!
It is strictly necessary to comply with
the max loads indicated (Kg) for each
shelf to avoid deformation or breakage.
7 Kg
7 Kg