Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker 1688-020-122 Handbuch Seite 6

4k-koppler, c-fassung, mit advanced imaging modality (erweitertes bildgebungsverfahren)
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 29
5. The coupler contains optomechanical parts that are precisely
aligned. Dropping the coupler can cause mechanical damage
and/or misalignment.
6. Test the coupler for functionality before using it in a surgical
procedure.
7. This coupler is shipped non-sterile. Clean and sterilize the coupler
prior to the first use and after every subsequent use. Follow the
cleaning, disinfection, and sterilization instructions provided in this
user manual.
8. Do not use the coupler and the camera system in the presence of
flammable or explosive gases or in an oxygen-rich environment.
9. No modification of this device is authorized by Stryker.
Operating the Endobody Clamp
To perform setup, disassembly, and cleaning of the coupler, the
endobody clamp (or "endoclamp") must be depressed and/or
released.
• To depress the endoclamp, push down as illustrated by the down
arrow.
• To release the endobody clamp, release pressure on the clamp
and allow the spring to push the clamp into its original position, as
illustrated by the up arrow.
Setting Up the Coupler
When attaching or removing the coupler, grip only the rear
adapter, as twisting other parts of the coupler with force can
result in mechanical damage.
Do not overtighten the coupler, as this can damage the
front window of the camera.
1. Gripping the rear adapter, screw the coupler clockwise onto the
camera head until it forms a tight seal.
2. Depress the endobody clamp
endobody
EN-2
.
2
and insert an endoscope into the
1

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis