Herunterladen Inhalt Inhalt Diese Seite drucken

Environmental Conditions - Chauvin Arnoux AmpFlex A110 Bedienungsanleitung

Flexible ac- stromwandler
Vorschau ausblenden Andere Handbücher für AmpFlex A110:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 46
The mean battery life is:
„ A110 single‑phase
300 h in continuous operation
1,800 10‑minute measurements
„ A130 three‑phase
500 h in continuous operation
3,000 10‑minute measurements
When the device is not connected to line power and the green On indicator goes off, the batteries must
be replaced (see § 4.2).
3.6.2. MAINS ADAPTER (OPTION)
The device can be powered by a standard mains adapter (5 V
connector.
When the adapter is connected and the switch is set to OFF, the On indicator is lit steadily.
When the device is on, the brightness of the On indicator varies to indicate that the batteries are OK. If
the indicator remains on steadily, the batteries must be replaced.

3.7. ENVIRONMENTAL CONDITIONS

The device must be used in the following conditions:
% RH
100
95
90
80
75
70
60
50
3
40
30
20
10
0
-40
-30 -20
1: Range of reference
2: Operating range
3: Storage range (without battery)
For an extended period of non‑use or storage, withdraw the batteries from the electronic unit.
The sensor by itself can withstand a maximum temperature of 90°C for 10 minutes.
Degree of pollution: 2.
Altitude: < 2000 m.
2
1
18
28
-10
0
10
20
35
55
30
40
50
60
39
, 100 mA) with a type B micro‑USB
dc
°C
70
80

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis