4.3.3 Environmental conditions

4.3.3 Environmental conditions
4.3.3 Environmental conditions
4.3.3 Environmental conditions
4.3.3 Environmental conditions
Indoor use
Altitude: 2000 m
Transportation altitude: 12000 m
Temperature influence: < .15% / 10°C
Climatic conditions
TEMPERATURE

4.4 CONFORMITY TO INTERNATIONAL STANDARDS

4.4.1 Electrical safety (in accordance with IEC 1010-2-032)
4.4.1 Electrical safety (in accordance with IEC 1010-2-032)
4.4.1 Electrical safety (in accordance with IEC 1010-2-032)
4.4.1 Electrical safety (in accordance with IEC 1010-2-032)
4.4.1 Electrical safety (in accordance with IEC 1010-2-032)
-
Dual insulation
-
Installation category III with operating voltage: 600 V
-
Installation category IV with operating voltage: 300 V
4.4.2 Electromagnetic compatibility
4.4.2 Electromagnetic compatibility
4.4.2 Electromagnetic compatibility
4.4.2 Electromagnetic compatibility
4.4.2 Electromagnetic compatibility
Emission: class B (i.a.w EN 50081-1)
Immunity (i.a.w EN 50082-2)
-
Electrostatic discharge (i.a.w IEC 1000-4-2):
4 kV in contact - level 2 - capability criterion B
8 kV in air - level 3 - capability criterion B
-
Radiated electromagnetic energy (i.a.w IEC 1000-4-3):
10 V/m - level 3 - capability criterion A
-
Short-duration transients (i.a.w IEC 1000-4-4):
2 kV - level 3 - capability criterion B
-
Line frequency magnetic field (i.a.w IEC 1000-4-8):
30 A/m - level 4 - capability criterion A
4.4.3 Mechanical protection
4.4.3 Mechanical protection
4.4.3 Mechanical protection
4.4.3 Mechanical protection
4.4.3 Mechanical protection
-
Tightness: protection level IP20 (i.a.w IEC 529) with jaws closed
-
Free fall: 1.5 m (i.a.w IEC 68-2-32)
-
Shocks: 100 g (i.a.w IEC 68-2-27)
-
Vibrations: .15 mm - 10/55/10 Hz (i.a.w IEC 68-2-6)
4.4.4 Flammability rating
4.4.4 Flammability rating
4.4.4 Flammability rating
4.4.4 Flammability rating
4.4.4 Flammability rating
-
VO (i.a.w UL 94)
I -
Reference domain
II - Operation domain
III - Storage domain
-
Pollution degree 2