Quick Setup

WARNING: It is essential to check settings during the process itself, and confirm
that they are correct, before regular operation commences.
Note: After adjusting values, download data to the device, then go to View – Display to track
the effects.
Initial setup can be carried out prior to mounting into the process, but it is extremely important
to calibrate the unit and adjust the sensitivity on the product itself.
Adjust 0 % application range setting
Application
General
applications
Demanding
applications
Interface
detection
CLS200 Digital
1.
2.
3.
CLS300 Digital
1.
1.
2.
7ML19985XA82
Downloaded from
www.Manualslib.com
Material
dry solids
low viscosity liquids
hygroscopic / wet solids
high viscosity and high
conductivity liquids
liquid A / liquid B
foam / liquid
Install the device in the process.
Open the menu View – Display and select the tab Transducer Block: Discrete Input
(Part 1). Make a note of the Sensor Value (digits) when the device is in the setup
condition that most closely represents the process (refer to table above).
Subtract 300 counts from the Sensor Value noted in step 2 and enter this new value in the
parameter view field: > Input > Transducer Block: Discrete Input > Range of Application>
0%.
Install the device in the process.
Open the menu View – Display and select the tab Transducer Block: Discrete Input
(Part 1). Make a note of the Sensor Value (digits) when the device is in the setup
condition that most closely represents the process (refer to table above).
Enter the sensor value in the parameter view field: > Input > Transducer Block: Discrete
Input > Range of Application> 0%.
Pointek CLS200/300 (Digital) – QUICK START MANUAL
manuals search engine
Setup conditions
Sensor uncovered and a minimum of 100 mm (4")
free space all around
Sensor immersed and then uncovered, but
retaining as much build up of material as possible
on the sensor.
Immerse the sensor in the material that has the
lowest dielectric constant.
Page EN-13