Controller MultiControl 2
Safety remarks
Use only in accordance with this manual. Store the manual in a space accessible to all users.
Must not be installed in a space that is subject to the danger of explosions.
Use only with rated voltage.
Only for use in laboratories. Operate only by trained personnel. Do not use in domestic
settings.
This autoclave must only be maintained by CertoClav or authorized retailers. Always use
original spare parts.
Do not open the case.
Do not install the controller close to high power electronics.
Install the controller only in a full regulating circle where the heat probe PT100 can gauge the
activity of the heater attached.
Do not install closer than 30cm to a CertoClav autoclave in order to protect the controller
from heat.
Use only appropriate material for wall mounting