Autoclave - operation
READ INSTRUCTIONS BEFORE FIRST USE!
The first run should be without load; only with water.
Loading the autoclave
Fill in 1.5 liters of distilled water. It is possible to insert more water, yet this will increase
heating up and cooling down time.
Put in instrument tray and basket and place loading. It is important that steam can circulate
freely.
Attention: Do not autoclave tightly sealed containers (especially glass bottles) since there is
no pressure compensation.
Close and Locking
Check if the gasket is inserted in the lid.
Press the red knob and turn it clockwise to OPEN.
Place the lid on the chamber and turn it clockwise until the handles are in line with
each other.
Turn the red knob counterclockwise to lock the lid.
We recommend connecting a tube to the steam release cock to direct the steam to a
condensation vessel. We offer both under art. no. 8582020.
Plug in the power cable. Plug in the heater cable into the pot and controller. Plug in
the temperature sensor cable and screw it into the controller.
ATTENTION: on Longtime cycles avoid cooking dry and overheating (could damage the
autoclave). Check the water level. The steam release cock has to be closed– remove the
seal for this.