Starting a program

Select a sterilization program in the control unit and push start.

Venting

If venting is activated, then the sterilization chamber is vented at 115°C for 10 minutes.

Sterilization / Autoclaving

For long time cycles close the cock. Remove the seal to enable this. This does NOT lead to a
loss of warranty. The seal is only for process quality control in sterilization processes where
the cock must not be closed.
Cooling down – end of cycle – acoustic signal
As the last step the controller cools down the autoclave to 80°C or 70°C for liquid program.
Thereafter an acoustic signal sounds and the controller displays "program finished".
The autoclave can now be opened.
If your application should require a faster cool down process, you can release the pressure
via the steam release cock. Obey safety remarks.

Opening

Turn red knob of locking to open = unlock. Steam release valve will be opened.
Steam escapes.
Turn lid anticlockwise and remove.
To get better drying of the loading remove the basket from the chamber and let it cool
down to room temperature before taking goods out.
Should the cycle interrupted caused by switch off or electrical failure or other reason –
the sterilization is not competed correct and has to be repeated. For opening please
follow the safety remarks for cooling down.