Opening and closing the lid
10.1
Opening the lid
The lid can only be opened when the centrifuge is switched on and the rotor is at rest. If it cannot be
opened under these circumstances, see the section on "Emergency release".
Press the button
OPEN / STOP
extinguishes.
The cover opens automatically approx. 45°.
This opening angle can be adjusted, if needed.
Turn the adjusting screw (A) with the hex head wrench included in delivery.
Turning clockwise:
Turning counterclockwise:
10.2
Closing the lid
Do not put your fingers between lid and housing.
Do not bang the lid shut.
If the left LED in the
lock goes into the basic position (opened).
Place the lid and lightly press down the front edge of the lid. The locking action is effected by motor. The left LED
in the button
OPEN / STOP
38/92
The lid unlocks via the motor and the left LED in the pushbutton
Cover is opened less.
Cover is opened wider.
button flashes, press the
OPEN / STOP
lights up.
Rev. 09 / 04.2020
button so that the motor-driven lid
OPEN / STOP
AB2400DEENFRIT
OPEN / STOP
A