Disposal of Equipment and Accessories

Follow your local governing ordinances for disposal and recycling of the Inogen One® G5 and
accessories. If WEEE regulations apply, do not dispose of in unsorted municipal waste. Within
Europe, contact the EU Authorized Representative for disposal instructions. The battery contains
lithium ion cells and should be recycled. The battery must not be incinerated.
Inogen One® G5 Pulse Volumes at Flow Settings
Breaths per
1
Minute
15
14
20
11
25
8
30
7
35
6
40
5
Total
Volume per
210
Minute (ml/
min)
Standards Compliance
This device is designed to conform to the following standards:
IEC 60601-1 Medical Electrical Equipment, Part 1: General Safety Requirements
IEC 60601-1-2 3.1 Edition, Medical Electrical Equipment, Part 1-2: General Safety
Requirements – Collateral Standard: Electromagnetic Compatibility; Requirements and
Tests
ISO 8359 Oxygen Concentrators for Medical Use – Safety Requirements. RTCA DO 160
Note: IT-network is a system composed of wireless (Bluetooth) transmission
between the Inogen One G5 and the Inogen Connect Application.
Connection of the Inogen One G5 to an IT-Network could result in previously unidentified
risks to patients, operators or third parties.
Subsequent changes to the IT-network could introduce new risks and require additional
analysis
Changes to the IT-network include:
º Changes in the IT-network configuration;
º Connection of additional items to the IT-network
º Disconnecting items from the IT-network
º Updating equipment connected to the IT-network
Inogen One® G5 Flow Setting
2
3
28
42
21
32
17
25
14
21
12
18
11
16
mL/breath +/- 15% per ISO 80601-2-67
420
630
4
5
56
70
42
53
34
42
28
35
24
30
21
26
840
1050
6
84
63
50
42
36
32
1260
24