Herunterladen Inhalt Inhalt Diese Seite drucken

Installation; Maintenance & Servicing; Intended Use; Equipment Disposal - VWR 5000IR Bedienungsanleitung

Inkubationsrundschüttler
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 64
I
nstallatIon
Upon receiving the VWR Incubating Orbital Shaker, check to ensure that no dam-
age has occurred during shipment. It is important that any damage that occurred in
transport is detected at the time of unpacking. If you do find such damage the carrier
must be notified immediately.
After unpacking, it requires two (2) people to lift the Incubating Orbital Shaker, from
the bottom, to place on a level bench or table, away from explosive vapors. Ensure
that the surface on which the unit is placed will withstand typical heat produced by the
unit. Always place the unit on a sturdy work surface.
The Incubating Orbital Shaker is supplied with a power cord that is inserted into the
IEC connector on the back of the unit first, then it can be plugged into a properly
grounded outlet. The 230V unit plugs into a 230 volt, 50/60 Hz source.
M
& s
aIntenance
ervIcIng
The Incubating Orbital Shaker is built for long, trouble-free, dependable service.No lubri-
cation or other technical user maintenance is required. However at least every three
(3) months you should:
• Unplug the unit.
• Remove any accumulated dirt from the base and tray.
• Check all accessible items to make sure they are properly tightened.
The unit should be given the care normally required for any electrical appliance. Avoid
wetting or unnecessary exposure to fumes. Spills should be removed promptly. DO
NOT use a cleaning agent or solvent on the front panel or lid which is abrasive or
harmful to plastics, nor one which is flammable. Always ensure the power is discon-
nected from the unit prior to any cleaning. If the unit ever requires service, contact your
VWR representative.
I
U
ntended
se
The Incubating Orbital Shaker is designed for a variety of shaking applications.
e
d
qUIPMent
IsPosal
This equipment must not be disposed of with unsorted waste. It is
your responsibility to correctly dispose of the equipment at life-cycle-
end by handing it over to an authorized facility for separate collection
and recycling. It is also your responsibility to decontaminate the
equipment in case of biological, chemical and/or radiological con-
tamination, so as to protect the persons involved in the disposal and
recycling of the equipment from health hazards.
For more information about where you can drop off your waste of equipment, please
contact your local dealer from whom you originally purchased this equipment. By doing
so, you will help to conserve natural and environmental resources and you will ensure
that your equipment is recycled in a manner that protects human health.
e
c
nvIronMental
ondItIons
Operating Conditions: Indoor use only.
Temperature:
15 to 32°C
Humidity:
maximum 80% relative humidity, non-condensing
Altitude:
0 to 2000 M above sea level
Non-Operating Storage:
Temperature:
-20 to 65°C
Humidity:
maximum 80% relative humidity, non-condensing
Installation Category II and Pollution Degree 2 in accordance with IEC 664.
EN
2

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

5000i

Inhaltsverzeichnis