Herunterladen Diese Seite drucken

Agilent Technologies Turbo-V 701 Bedienungshandbuch Seite 260

Rack controller

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 27
17
Technical Information
Programming
This menu contains all the parameters that have a direct effect on
driving of the pump.
HIGH SPEED ADJ: Sets the rotational speed of the pump when the
low-speed function is not active, expressed either in Hz or KRPM.
LOW SPEED ADJ: Sets the rotational speed of the pump when the
low-speed function is active, expressed either in Hz or in KRPM.
SOFT START: enables (YES) or disables (NO) the soft start function.
This parameter is visible only if the controller is configured in
FRONT mode.
EXTERNAL FAN: it's possible to operate the external cooling fan
selecting between following options:
VENT VALVE MODE: it's possible to operate the optional vent valve
selecting between following mode of operation:
VENT VALVE DELAY: Sets the delay, expressed in seconds, between
stopping of the pump and opening of the vent valve.
VENT OPEN TIME: sets the time, expressed in seconds, for the vent
vale to stay in open condition. Enter 0 if the valve has to stay open
until the next start.
LOAD GAS TYPE: selects the type of gas used in the process.
Therefore, the controller limits the power supplied to the pump
according to this setting.
260/288
ALWAYS ON: the cooling fan is always ON
OFF_LIMPOWER: 10 seconds after the pump has
reached NORMAL status, the fan is switched off and
the driving power of the pump is reduced to XXX.
SERIAL: the cooling fan is controlled directly via the
serial port.
AUTO TIME: the valve is controlled by the controller
SERIAL: the valve is controlled by the serial port.
AUTO PUMP SPEED: the valve is controlled by the
controller its king into account the pump's
deceleration.
Turbo-V 701 Rack Controller User Manual / 87-901-032-01

Werbung

loading

Diese Anleitung auch für:

X3501-64002