Herunterladen Diese Seite drucken

Agilent Technologies Turbo-V 701 Bedienungshandbuch Seite 234

Rack controller

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 27
17
Technical Information
RS 232/RS 485 Communication Description
Tab. 5
Data Type
Logic (L)
Numeric (N)
Alphanumeric (A)
The addressed SLAVE will respond with an ANSWER whose structure
depends from the MESSAGE type.
When the MESSAGE is a reading command, the SLAVE will respond
transmitting a string with the same structure of the MESSAGE.
Using the RS 485 interface, the message structure remains identical to the one used for
NOTE
the RS 232 interface, the only difference being that the value assigned to the ADDRESS
<ADDR>.
234/288
<COM> (Command) = 0x30 to read the window, 0x31 to write into
the window
<DATA> = an alphanumeric ASCII string with the data to be
written into the window. In case of a reading command this field
is not present.
The field length is variable according to the data type as per the
following table:
<ETX> (End of transmission) = 0x03
<CRC> = XOR of all characters subsequent to <STX> and
including the <ETX> terminator. The value is hexadecimal coded
and indicated by two ASCII character.
Turbo-V 701 Rack Controller User Manual / 87-901-032-01
Field Length
Valid
Characters
1
'0' = OFF
'1' = ON
6
'-', '.', '0' . . . '9' right
justified with '0'
10
from blank to '_' (ASCII)

Werbung

loading

Diese Anleitung auch für:

X3501-64002