Instructions for Use
Error Messages
MESSAGE
FAULT: TOO HIGH
LOAD
FAULT: SHORT
CIRCUIT
OVERVOLTAGE
206/288
DESCRIPTION
In normal operation, the current
drawn by the pump is higher
than programmed.
During normal functioning (after
the start-up phase), a short-
circuit condition has been
detected in the power circuit.
A failure has occurred in the
controller power supply section,
or the controller has received a
disturbance from the mains.
Turbo-V 701 Rack Controller User Manual / 87-901-032-01
REPAIR ACTION
Check that the pump rotor
is free to rotate.
Press the START push-
button twice to start the
pump.
Check connections and
shortages between pump
and controller. Press the
START push-button twice
to start the pump.
Press the START button
twice to start the pump
again. If the message is
redisplayed, contact
Agilent for maintenance.