7.5 Seals and Tubes

To insure proper function of the Suction Pump, all tubes and seals must be periodically checked,
and replaced after at most 250 sterilizations or five years operation. Only when in perfect order can
sufficient vacuum be built up.
Defective seals and tubes must be replaced!
2
1. Seal for connecting neck on Control Unit......................................................... 2 pieces ....... (Art. 4063)
2. Bacteria filter tube between Bacteria Filter and Jar lid, 0.5 m............................ 1 piece ....... (Art. 4190)
3. Seal for unscrew able connecting neck on Secretion Jar Lid .......................... 4 pieces ....... (Art. 4064)
4. Seal between Jar Lid and Jar............................................................................. 1 piece ..... (Art. 28957)
5. Seal between wide unscrew able connector neck and Secretion Jar Lid........ 4 pieces ....... (Art. 4056)
6. Seal in Overflow Protection ................................................................................ 1 piece ..... (Art. 28958)

7.6 Control Unit Fuse Replacement

A blown Fuse can be easily replaced by the operator. It is located on the back of the Control Unit:
1)
Disconnect mains.
2)
With screwdriver open fuse cover.
3)
Replace defective fuse. (see 1.2 Technical Data)
4)
Close fuse case.
5)
Reconnect mains.

7.7 Disposal

Disposal of device, components and accessories must strictly conform to local laws and regulations
as set out by the relevant authorities. With regard to the preservation of the environment old
equipment may be returned to the distributor or manufacturer.