6.5 Secretion Jars with bottle lids.

After every use, Secretion Jars with bottle lids are to be emptied, rinsed and disinfected before
autoclaving at maximum 135°C.
After Autoclaving, allow Secretion Jars to dry for one hour at room temperature with the paper side up.

6.6 Bacterial Filter, hydrophobic

The Bacteria-filter between the Secretion Jar and Control Unit Pump is a one-
way device and Cannot be cleaned. It should be replaced periodically every
~8 h, or immediately should foam or bacterial substance appear.
In order to protect the pump, the Bacteria filter closes when it comes in contact
with liquid. So further suction procedure is not possible when the Bacterial filter is
saturated.

6.7 Bacteria filter tube

1)
Rinse Bacteria filter tube well with disinfecting agent.
2)
Clean with damp cloth.
3)
Bag Bacteria filter tube and autoclave at maximum 135°C.
To insure proper function of the Suction Pump, all tubes and seals must be periodically checked, and
replaced after at most 250 sterilizations or five years operation.
The One-way Tubing Set No. 6017 and No. 6026 cannot be sterilized!

7.1 Technical Safety Inspections

Various countries require technical safety inspections of medical devices in regulations. The technical
safety inspection is a legal periodic safety check required for operators of medical devices. The
objective of this measure is the timely identification of faults and risks for patients, users or third
parties.
The Inspection interval for the Vacuson 60LP was stipulated on 2 years (based on Appendix 1 of the
MPBetreibV) [German Medical Devices Operator Ordinance]:
NOUVAG AG offers technical safety inspections for clients. Please contact our technical customer
service.
No. 31793 15/11

7 Servicing

Vacuson 60LP
13