WARNING!
Before replace Tip-Seals remove the power supply cable.

To Remove the Worn Tip-Seals

Disconnect the TPS-compact from electrical power.
Unlock the scroll pump support by unscrewing the screws (A) (see
the "Tip-Seal replacement - Figure A")
1
2
3
4
TPS-compact User Manual / 87-901-010-01 (H)
Lift the TPS-compact in order to access the Scroll pump site
(you'll hear a "click" noise as soon as the upper part of TPS-
compact is locked).
Remove (4) M5 socket head bolts (B).
Remove the outboard housing axially off the frame (C).
Remove and discard the worn tip seals (D) and the main O-ring.