Test Preparation:
Use deionized (DI) water to thoroughly clean the PosiTector SST
test cell, test cup, syringe and PosiPatch (if used in test).
All test materials should be handled in a manner that minimizes the
potential for contamination. The PosiTector SST operator may
choose to wear nitrile or other non-contaminating gloves when
performing the test.
Remove the protective cap from the probe. Power-up the Gage by
pressing the center navigation
Mode, the Gage guides you through three simple steps - 1, 2 and 3.
Step 1: Obtain Background (Blank) Conductivity Measurement (
NOTE: The Gage automatically displays the previous background
measurement, if any. This measurement can be reused and step (c)
to step (e) can be skipped by pressing the (-) button.
a) Pour at least 4ml (milliliters) of deionized (DI) water into the
supplied cup (to prevent contamination of the water source).
b) Fill the syringe completely with DI water from the cup (slightly
more than 3ml).
c) Completely fill the PosiTector SST probe's test cell with water
from the syringe (approximately 1ml).
d) Wait until the measurement stabilizes (up to 5 seconds), then
press the (+) button to save a new background conductivity
measurement (
e) Use the syringe to draw all of the water from the test cell. Shake
any remaining water out of the test cell.
f) With the syringe pointing up, expel the air and excess solution
from the syringe until there is 3.0ml of DI water remaining.