Herunterladen Inhalt Inhalt Diese Seite drucken

Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Gima KD-558 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für KD-558:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 93
Table 2
For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer's declaration - electromagnetic immunity

The [KD-558] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [KD-558] should assure that it is used in such
an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
ENGLISH
IEC 60601
Compliance
test level
± 6 kV
contact
± 8 kV air
3 A/m
level
environment - guidance
± 6 kV
Floors should be wood,
contact
concrete or ceramic tile. If floors
± 8 kV air
are covered with synthetic
material, the relative humidity
should be at least 30%
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
Electromagnetic
38

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

32921

Inhaltsverzeichnis