39
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not
LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The [KD-558] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [KD-558] should assure that it is used in such
an environment.
IMMUNITY test
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
IEC 60601 Compliance
test level
level
3 V/m 80
3 V/m
MHz to
2.5 GHz
ENGLISH
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
[KD-558], including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency
of the transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2,5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
recommended separation distance
in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
be less than the compliance level
in each frequency range.
Interference may occur in the vicinity
of equipment marked with
the following symbol:
should
a
b

Quicklinks ausblenden: