Cleaning:
Manual
Disinfection:
Manual (Inter-
mediate-Level)
Packaging
Sterilization
Drying
Maintenance
Storage
Manufacturer
Contact
1. The SmartLite
®
have to be cleaned by scrubbing with a pre-impregnated wipe
or disposable towel soaked with an alcohol-based, tubercu-
locidal, quaternary ammonium solution with a label claim for
cleaning (e.g., VoloWipes disinfecting  /   c leaning  /   d eodorizing
Wipes) or pH-neutral, phosphate-free cleaning solution (e.g.,
Dr. Schumacher Instru Plus [3%]) according to local regula-
tions and use according to cleaning solution manufacturer's
Directions for Use until it is free of visible residues. Remove all
visible soil, ensuring fluid penetrates all crevices. Do not allow
solution to penetrate the casing. Discard used wipes. Additional
wipes may be used.
2. Allow the test article to air dry.
1. After cleaning wipe all device surfaces with a new single use
cloth in combination with an alcohol-based, tuberculocidal, qua-
ternary ammonium solution e.g.: VoloWipes disinfecting  / clean-
ing  /   d eodorizing Wipes, BBraun Meliseptol
according to local regulations, and use according to disinfectant
solution manufacturer's Directions for Use. Pay special attention
to seams, areas around buttons, window, and crevices.
2. Allow the test article to air dry.
3. Wipe the test articles with a sterile, clean, lint-free cloth damp-
ened with distilled water.
4. Pay special attention to device seams and insertions.
No particular requirements.
Sterilization is not allowed. No methods have been validated.
Do not subject components to Steam autoclaving or liquid chemi-
cal sterilant immersion. Component damage will occur.
Wipe the test articles dry with a sterile, clean, lint-free cloth. Allow
the components to fully air dry before storage.
Visually inspect to ensure that all contamination has been re-
moved.
Visually inspect power supply and cord for damage.
Components that are damaged, worn, or distorted should be dis-
carded.
No additional maintenance or lubrication is recommended.
Store the SmartLite
at room temperature, away from moisture or excessive humidity.
Within the United States, call Dentsply Sirona at 1-302-422-4511.
For areas outside the United States, contact your local Dentsply
Sirona representative.
Focus
light handpiece and charging base
®
Focus
light handpiece and charging base
®
®
Rapid, approved
®
9

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis