Mechanical installation

4.2

Communication interfaces

®
4.2.1 Bluetooth
interface
After the start sequence, the Bluetooth
detector is activated.
If no connection is established, the gas detector disables the
interface after 10 minutes (factory setting).
If the connection is to be secured with a pairing PIN, an
additional pairing can be carried out and an individual pairing
PIN assigned. The factory setting is the connection without
PIN.
For safety-related applications, an individual pairing PIN
must be assigned.
4.2.2 Activating the Bluetooth
®
● Activate the Bluetooth
via the ISA100 Wireless
● Magnetic pen: Tap the marking (Fig. A 12).
● ISA100 Wireless
: Activate using PolySoft.
✓ The blue LED flashes.
4.2.3 Establish Bluetooth
pairing
Prerequisite:
®
– Bluetooth
interface on the device and on the Windows
terminal device is activated.
● Connect via PolySoft.
Entering a PIN is not necessary.
✓ After successful connection, the blue LED lights up
continuously.
Gas detector and Windowsterminal device with PolySoft
are connected.
4.2.4 Carry out Bluetooth
Prerequisite:
®
– Bluetooth
interface on the device and on the Windows
terminal device is activated.
®
– Bluetooth
pairing is activated (with PolySoft).
®
● If the Bluetooth
pairing PIN "000000" is used, Bluetooth
pairing is automatic.
If an individual Bluetooth
a prompt requests entering the PIN.
✓ After successful connection, the blue LED lights up
continuously.
Gas detector and Windowsterminal device with PolySoft
are connected.
If required, use PolySoft to assign an individual Bluetooth
pairing PIN.
NOTICE
► Safeguard the individual Bluetooth
26
®
interface of the gas
®
interface
interface with a magnetic pen or
network:
®
connection without
®
pairing and connect
®
pairing PIN has been assigned,
®
pairing PIN.
Instructions for use
4.2.5 Connecting the ISA100 Wireless
interface
● Connect the Windows mobile device with the respective
management station of the ISA100 Wireless
The configuration is done with PolySoft.
5
Mechanical installation
WARNING
Risk of explosion!
In potentially explosive atmospheres, electrostatic discharge
may cause sparks on the antenna.
► Do not install the antenna in explosion-hazard areas
subject to intense electrostatic charging, for example, due
to fast moving particles or pneumatic transport of
powders.
WARNING
Risk of explosion!
In areas where there is a risk of dust explosion, electrostatic
discharge may cause sparks on the housing and sensing
head.
► Do not install the gas detector in dust explosion-hazard
areas subject to intense electrostatic charging, for
example, due to fast moving particles or pneumatic
transport of powders.
5.1
Installing the gas detector -
installation without mounting
accessories
The gas detector is installed vertically on a flat surface.
For installation with mounting accessories, see
corresponding assembly instructions.
Prerequisites:
– A suitable installation location is selected (see "Safety
information", section "Mechanical installation", page 24).
®
CAUTION
Risk of explosion!
The gas detector is not tested in environments with increased
oxygen content (> 21 % O
► Do not use the gas detector in environments with
increased oxygen content.
Install the gas detector:
®
● Install the gas detector using the drilling template (see
"Drilling template" appendix) and M6 (1/4") hexagon
socket head cap screws.
5.2

Installing the antenna

1. Remove the cover cap from the antenna connector
(Fig. A 4).
®
|
Dräger Polytron
6100 EC WL, Dräger Polytron
network.
).