Carrying out the span calibration:
1. Start and confirm span calibration in PolySoft.
 ISA100 Wireless
value status "Bad," (substatus maintenance mode
active).
2. Follow the instructions of the wizard In PolySoft.
Flow with test gas supply: 0.5 L/min ± 10 %
Display
Access
PV
READ
PV
READ
PV
READ
PV
READ
PV
READ
DIAG_STATUS
READ
DIAG_STATUS
READ
DIAG_STATUS
READ
DIAG_STATUS
READ
11.2 Measured value status(PV_STATUS)
Meaning of "Quality" (Bit 6 and 7):
0: Bad (Bad) – No measured value transmission.
1: Uncertain (Uncertain) – Measured value quality lower than normal.
2: Good (Good) – Measured value quality good, but alarm condition could be present.
Quality
Substatus
Bits 6 and 7 Bits 2 to 4
0
0
0
1
0
2
0
3
0
4
Instructions for use
|
Dräger Polytron
interface transmits the measured
Byte no.
1
2
3
4
5
1
2
3
4
PV
Cause
NaN
Start sequence active.
Start sequence failed, no EC sens-
ing head detected.
NaN
Sensor lock feature is enabled. A
sensor with a different part number
was inserted.
NaN
Gas detector not connected to
ISA100 Wireless
by System Manager). Possible
causes:
– ISA100 Wireless
can not be found
– ISA100 Wireless
settings changed
NaN
Serious device fault, various
causes.
NaN
Sensor not detected or defective.
®
6100 EC WL, Dräger Polytron
3. After completion of the span calibration, switch off the gas
flow and remove the calibration adapter.
11

Troubleshooting

11.1 ISA100 Wireless
Content
Modbus input
register
PV_Status
Register N
PV_Value (1/4)
Register N+1
PV_Value (2/4)
PV_Value (3/4)
Register N+2
PV_Value (4/4)
Diag_Status (1/4)
Register M
Diag_Status (2/4)
Diag_Status (3/4)
Register M+1
Diag_Status (4/4)
Remedy
Wait for start sequence.
A Restart the gas detector.
B Have the gas detector checked by Dräger-
Service.
Use a sensor with the same part number or dis-
able the sensor lock feature.
A Check ISA100 Wireless
network (used
B Check ISA100 Wireless
C Check data for integration in
ISA100 Wireless
network
network.
network
A Restart the gas detector.
B Reset the gas detector to factory settings.
C Update firmware.
D Have the gas detector checked by Dräger-
Service.
A Check the sensor connection.
B Remove and reinstall the sensor.
C Change the sensor.
®
Repeater ISA
Troubleshooting
standard objects
Data type
UINT (16 bit)
32-bit float (4 bytes), Float32
acc. to IEEE 754
UINT (16 bit)
UINT (16 bit)
receipt.
network settings.
|
en
31