Herunterladen Inhalt Inhalt Diese Seite drucken

M400 Type 1 Cond Ind Only: Conductivity Calibration For Inductive Sensors - Mettler Toledo M400 Kurzanleitung

Vorschau ausblenden Andere Handbücher für M400:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 19
CO
Process Calibration
2
Start as in chapter CO
2
A
17.3
hPa
A
27.3
°C
CO2 Calibration
Type = Process
A
17.3
hPa
A
27.3
°C
A Point1 = 00000 hPa
A
CO2 = 17.3
hPa
A
17.3
hPa
A
27.3
°C
A Point1 = 16.90 hPa
A
CO2 = 17.3
hPa
A
17.3
hPa
A
27.3
°C
pH S=100.0%
Z=7.009pH
Save Adjust
9

M400 Type 1 Cond Ind only: Conductivity Calibration for inductive Sensors

(PATH: Cal)
For Display setup, analog outputs and setpoints see the section "General Setup".
Cond Ind Zero point Calibration
A
1.25
µS/cm
A
25.00
°C
Calibrate Sensor
Channel A Conductivity
One point Calibration described.
Select Process calibration.
Take a sample and press the [ENTER] key again to store the current measuring value.
To show the ongoing calibration process, A or B (depending on the channel) is blink-
ing in the display. After determining the CO
to proceed with the calibration.
Enter the CO
value of the sample then press the [ENTER] key to start calibration.
2
The display shows the slope calibration factor S and the offset calibration factor Z.
After a successful calibration, the calibration values are taken over (Adjust) or were
aborted (Calibrate or Abort).
While in measurement mode press the key [CAL]. Press the
Conductivity calibration and press the [ENTER] key to call the sub function.
value of the sample, press the
2
or
key again
key to select the
15

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis