10 Conductivity Quick Setup
For Display setup, analog outputs and setpoints see the section "General Setup".
Sensor Type Selection
1.25
µS/cm
25.0
°C
Sensor Type = Cond(2)
Cell Constant
1.25
µS/cm
25.0
°C
p M=0.1003
A=0.0000
s M=1.0000
A=0.0000
Measurement units
1.25
µS/cm
25.0
°C
a S/cm
Analog Output?
Yes
11 Conductivity Calibration for 2-e and 4-e Sensors
For Display setup, analog outputs and setpoints see the section "General Setup".
Cond One point Calibration
A
1.25
µS/cm
A
25.0
°C
Calibrate Sensor
Channel A Conductivity
(PATH: Menu / Quick Setup)
Select the type of sensor to be used with the M400 transmitter. Choices are "Cond(2)",
used for all 2-Electrode type sensors and "Cond (4)" for all 4-electrode sensors. Press
the [ENTER] key.
Enter the appropriate cell constant(s):from the sensor label or certificate (M) for 2-elec-
trode sensors, leaving (A) at 0.000; or (M) and (A) values for 4-electrode sensors.
Press the [ENTER] key.
Select the measurement (conductivity or temperature) and units for measurement.
If using analog output, select Yes. Refer back to section "General Setup" to continue
setup.
While in Measurement mode press the [CAL] key. If the display prompts you to enter
the calibration security code, press the
then press the [ENTER] key to confirm the calibration security code.
Press [ENTER] to select the Conductivity calibration sub function.
A flashing "H" (alternating with "A" or "B" to show the channel being calibrated for
multi-channel devices) in the top left hand corner shows the ongoing calibration process.
(PATH: Cal)
or
key to set the calibration security code,
17